2) To recall data from memory, press . The location of the most recent stored data is displayed first. Press or to select the location of the desi
Note: When multiple slopes exist, the slope that is available to view will depend on the measured value before entering SETUP mode. For example, if p
5.6 P3.3 buFF (pH Buffers & Calibration Points) Press . Press or to select “USA” or “NIST” buffer group (pH mode only). Press or to select
6. TROUBLESHOOTING GUIDE PROBLEM CAUSE SOLUTION No display a) Main power not switched on. b) AC Adapter socket not inserted properly. a) Switch on
CAUSE SOLUTION From any screen Invalid key; Button does not work in the current operation mode. Press alternate key. Select valid buttons depen
13 7. ION SELECTIVE ELECTRODES Contact our partners for more details regarding ISE and related solutions. The table below lists the most common ele
14 8. SPECIFICATIONS pH Range 0.00 to 14.00 pH Resolution 0.01 pH Accuracy 0.01 pH + 1 count Calibration Points Up to 5 points wit
15 9. WARRANTY This meter is supplied with a warranty against significant deviations in material and workmanship for a period of THREE years from
For more information on our products, please contact our channel partner or visit our websites listed below: Oakton Instruments 625 E Bunke
TABLE OF CONTENTS 1. INTRODUCTION ... 1 2. GETTING STARTED...
1 1. INTRODUCTION Thank you for purchasing our pH 700 / Ion 700 series benchtop meter. These microprocessor-based meters are economical and simple
2. GETTING STARTED 2.1 Keypad Functions Powers the meter on and off. Upon power on, the meter automatically begins in the mode that was last used
2.2 LCD Annunciators 2.3 Meter Connections pH BNC connection for pH, ORP (Redox), or Ion Selective Electrodes (ISE). REF Pin connection for half
3. CALIBRATION 3.1 pH Calibration For best results, periodic calibration with known accurate standards is recommended. Calibrate with standards tha
4) To calibrate another buffer repeat step 3) or press to return to the measurement mode. Note: The meter will automatically return to measurement m
3.3 Millivolt (mV) Offset Adjustment Oxidization Reduction Potential (ORP or Redox) is not a precise measurement, but is useful as a relative indicat
1) Connect the ISE and press as needed to select ppm. Note: The primary reading will show “- - - -” without a stored ion calibration. The secondary